End of Run Shutdown

At the end of the last night of IGRINS observing, the following steps must be completed before IGRINS is removed from the telescope.
  1. Home the calibration motors:
    1. Open a terminal window and type “cd ~/igrins_admin”
    2. Type “python mtini.py”, it will ask to close HKP. 
    3. Follow the commands of that program to home the motors (select "0" to initialize then "5" to exit).

  2. Detector Shutdown:
    1. Using “Remote Desktop Connection” log onto each of the three detector computers to shutdown the detectors and Jade2 cards.
      1. IP Addresses for each computer are listed here:
        1. SVC: 192.168.2.4
        2. H: 192.168.2.5
        3. K: 192.168.2.6
      2. When logged into Windows navigate to the “HXRG Testing Software” window.
      3. In the top left panel, press the “RESET ASIC” button. This will grey out most of the other options on the panel.
      4. In the top left panel, uncheck the “ASIC Main Power” box to power down the detector.
      5. Close the “HXRG Testing Software” window
      6. Navigate to the “HalServiceApp” window.
      7. Press the “Stop” button.
      8. Close the “HalServiceApp” window.
      9. Using the Windows Start menu shut down the computer. Answer the question regarding shutdown.

  3. At this point IGRINS can be safely removed from the telescope. All other IGRINS hardware can be shutoff at the power strip.

  4. Leave the ICC and DSC running. Close all terminals and start the TeamViewer software on each system.

  5. Unplug the DSC display cable from the monitor to avoid confusion when other instruments are in use. Tuck ICC and DSC keyboards and mice back and out of easy reach.