Startup:

1. North Control Room:

  • Servo amplifiers: 'Declination' and 'Hour Angle' buttons

  • Hydraulic bearing: 'Start' button

2. Dome Floor:

  • IMPORTANT: Ensure weather conditions are within acceptable limits before proceeding.

  • Unplug the extension cord from the HEPA filter.

  • Press the green 'Reset' button located on the Telescope Console.

  • Pin the doors open.

  • Open louvers via the panel located near the doors on the east wall with ‘LL Open.’

  • On the console, depress the 'Dome Open' button until the shutter is fully opened.

Note: The console does NOT correctly display the state of the shutter. Rely on visual and audio cues to indicate when the shutter is fully opened.

  • On the Hand-Paddle, select 'Upper Windscreen' and move it North until it stops.

  • Open the mirror cover via the console.

  • Extend the light shield.

Note: For GCMS (VIRUS-P) and VIRUS-W ONLY (Generally any instrument at f/9 will use the light shield / baffle).

  • In the TCS computer, select Tools → Enable Dome Automation

3. Slit Room (for TS2 ONLY):

  • Slew M5 turret to Slit #2 Silver.

  • Carefully uncover M5.

  • If slit cover is present, remove it carefully.

  • Uncover Guide Reflector and Guide Camera if covered.

Shutdown:

1. Dome Floor:

  • Retract the light shield if extended.

  • Close mirror cover.

  • Close dome shutter.

  • In TCS: Next → Stow → Yes → Go Next.

  • Slew to stow.

  • Center the upper windscreen over the telescope.

  • Close doors and louvers.

  • Turn on Passage Lights.

  • E-stop the console.

2. Slit Room (for TS2 ONLY):

  • Carefully cover M5.

  • No labels